•3 Hz to > 3000 MHz in two overlapping ranges
•High input sensitivity over the full frequency range
•High impedance measurement up to 125 MHz
•Reciprocal measurement technique gives superior resolution
•Period measurement from 3 Hz to 125 MHz
•Selectable measurement time; display hold function
•Noise filter for low frequency measurements
•Large 8.5 digit display with full range of annunciators
•Battery operation; handheld format with tilt-stand
•Push-to-measure function with auto power-down
Specifications:
1-Input:
-Input A:
•Input Impedance: 1MW//20pF (AC coupled)
•Frequency Range: <3Hz to >125MHz
•Sensitivity: Sinewave - 15mVrms 10Hz-125MHz
-Input B:
•Input Impedance: 50W (AC coupled)
•Frequency Range: <80MHz to >3000MHz
•Sensitivity: Sinewave - 15mVrms 80MHz-2GHz, 50mVrms to 3GHz
-Maximum Input Voltage: Input A and Input B: 30Vdc; 30Vrms 50Hz/60Hz reducing to 1Vrms above 1MHz
2-Measurements:
-Frequency (Range A):
•Frequency Range: 3Hz to 125MHz
•Resolution: 10-7Hz to 100Hz (see Note)
-Frequency (Range B):
•Frequency Range: 80MHz to >3000MHz
•Resolution: 1Hz to 10kHz (see Note)
-Period (Range A only):
•Period Range: 8ns to 330ms
•Resolution: 10-7ns to 1us (see Note)
-Measurement Time: Selectable as 10s, 1s or 0.1s. (Note that for the 0.1s setting the effective measurement time is 0.3 seconds).
-Resolution: The displayed resolution depends upon the measurement time and the input frequency. Eight digits (or nine using an overflow indication) are displayed for a 10s measurement time. Seven or eight digits are displayed for a 1s measurement time, and six or seven digits for 0.1s depending upon the input frequency. Usable resolution may be further reduced by noise, particularly at low frequencies.
-Accuracy: Measurement accuracy is the sum of the timebase accuracy and measurement resolution plus
one count.