The Pfeiffer Vacuum IMR 265 is a compact process ion gauge designed for accurate vacuum measurement across a wide pressure range. By combining Pirani and hot cathode sensors, it ensures precise monitoring from high to ultra-high vacuum levels. The device intelligently controls the activation of the hot cathode sensor based on pressure thresholds to extend filament life and enhance measurement reliability. Its corrosion-resistant construction and compatibility with various control units make it a versatile choice for diverse vacuum applications.​

Key Features:

  • Dual Sensor Design: Integrates both Pirani and hot cathode sensors within a single compact unit, providing a broad measurement range from 2Ɨ10⁻⁶ hPa to 1000 hPa.​
  • Automatic Sensor Control: The hot cathode sensor is activated only below a user-defined pressure threshold (factory set at 5Ɨ10⁻² hPa) to prevent filament burnout, while the Pirani sensor monitors higher pressures.​
  • High Measurement Accuracy: Offers precise pressure readings across its entire range, suitable for various vacuum applications.​
  • Corrosion-Resistant Construction: Built with materials resistant to corrosion, ensuring durability and reliability in demanding environments.​
  • Compact and Versatile: Its design allows for easy integration into existing systems, compatible with Pfeiffer Vacuum measurement and control units or other control devices.​
  • User-Adjustable Thresholds: Features five definable switching thresholds for the hot cathode activation, allowing customization based on specific process requirements.​

Additional Specifications:

  • Measurement Range: 2Ɨ10⁻⁶ hPa to 1000 hPa
  • Sensor Type: Dual Sensor (Pirani + Hot Cathode Ionization)
  • Flange Type: DN 25 ISO-KF
  • Output Signal: 0 to 10.2 V analog output
  • Supply Voltage: 20 to 30 VDC
  • Max Operating Pressure: 500 kPa absolute
  • Operating Temperature Range: 0 °C to +50 °C
  • Materials in Vacuum Path: Stainless Steel, Molybdenum, Iridium, Glass, Tungsten
Operating Voltage 20 - 30V
Sensor Type Gas Sensor
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Stock

Shipping Method

ETA

Cost

Available

Self Pickup

1 Day

Free

Available

inside UAE

2 - 3 Days

22 AED / Free Above 50 AED

Available

International

4 - 7 Days

180 AED / 49 $

pre-order

General

4-5 Weeks

 

Pfeiffer Vacuum IMR 265, Dual Sensor (Pirani and Hot Cathode), DN 25 ISO-KF Flange

AEDĀ 8,950.00